Cleaning level :HEPA:ISO 5Level(100Level) ULPA:ISO 4Level(100Level)
Air supply air volume:650
Exhaust air volume:1100
Descend wind speed≥0.35m/s
Wind speed in≥0.55m/s
Noise≤58dB(A)
Half ventilation peak value≤3µm
Biological safety: Personnel protection: impact type sampler: CFU≤10; narrow slot type sampler: CFU≤5
Product protection: all sampling vessels: CFU≤5
Cross contamination protection: all double dishes: CFU≤2
Working area: 1100*600*640
Overall dimension1300*815*2210
Specification of air-supply filter1015*460*70*①
Number: Single
Biological Safety Cabinets